Universal RTD Calibrator

CL515-PLUS
Download PDFs:
PDF Icon

The CL515-PLUS works with the RTDs you use including Platinum 100 (alpha = 3850, 3902, 3926) & 1000 (alpha = 3850, 3750) Ω, Copper 10 & 50 Ω, Nickel 100 and 120 Ω. Easily set any value quickly to within 0.1° with the adjustable digital potentiometer “DIAL” plus store any three temperatures for instant recall with the check switch.

Calibrate quickly with automatic output stepping, choose between 2, 3, 5, 11 and 21 steps to automatically increment the output in 100%, 50%, 25%, 10% or 5% of span. Select the step time to match your system from 5, 6, 7, 8, 10, 15, 20, 25, 30 and 60 seconds.

Carry it without worry, it comes protected with a rubber boot and rugged, low profile switch. Easy to operate even in dark areas with the backlit display.

Specifications (Unless otherwise indicated all specifications are rated from a nominal 23°C, 70% RH for 1 year from calibration)

  • Accuracy: ±(0.015% of Reading + 0.05 Ω)
  • Temperature Drift: ± 0.01% of span outside of 23°C ±10°C (73°C ±18°F) Operating Temperature
  • Range: -25 to 60°C (-10 to 140°F)
  • Relative Humidity Range: 10% ≤RH ≤90% (0 to 35°C), non-condensing, 10% ≤RH ≤70% (35 to 60°C), non-condensing
  • Dimensions: 143 L x 76 W x 41 mmH (5.63 x 3.00 x 1.60″)
  • Weight: 0.34 kg (12.1 oz) including boot and batteries
  • Power: 4 “AA” Alkaline 1.5V (LR6) (included)
  • Battery Life: 50 Hours low Battery low battery indication with nominal 1 hour of operation left. Protection against misconnection over-voltage protection to 60 Vdc (rated for 30 seconds)
  • Display: High contrast graphic liquid crystal display. LED backlighting for use in low lit areas.

Indicate

  • Excitation: Current 0.5 mA nominal
  • Normal Mode Rejection: 50/60 Hz, 50 dB
  • Common Mode Rejection: 50/60 Hz, 120 dB

Source

  • Accuracy: From 1 to 10.2 mA external excitation Current ± (0.015% of Reading + 0.05 Ω)
  • Below: 1 mA of external excitation
  • Current: ± (0.015% of reading + 0.025 mV/mA excitation current + 0.05 Ω)
  • Resistance Ranges: 0.00 to 410.00, 410.1 to 4001.0 Ω
  • Allowable Excitation Current Range <410 Ωs: 10.2 mA max; steady or pulsed/intermittent
  • Allowable Excitation Current Range 410 to 4001 Ωs: 1 mA max; steady or pulsed/intermittent
  • Pulsed Excitation Current compatibility: DC to 0.01 second pulse width 

The CL515-PLUS works with the RTDs you use including Platinum 100 (alpha = 3850, 3902, 3926) & 1000 (alpha = 3850, 3750) Ω, Copper 10 & 50 Ω, Nickel 100 and 120 Ω. Easily set any value quickly to within 0.1° with the adjustable digital potentiometer “DIAL” plus store any three temperatures for instant recall with the check switch.

Calibrate quickly with automatic output stepping, choose between 2, 3, 5, 11 and 21 steps to automatically increment the output in 100%, 50%, 25%, 10% or 5% of span. Select the step time to match your system from 5, 6, 7, 8, 10, 15, 20, 25, 30 and 60 seconds.

Carry it without worry, it comes protected with a rubber boot and rugged, low profile switch. Easy to operate even in dark areas with the backlit display.

Specifications (Unless otherwise indicated all specifications are rated from a nominal 23°C, 70% RH for 1 year from calibration)

  • Accuracy: ±(0.015% of Reading + 0.05 Ω)
  • Temperature Drift: ± 0.01% of span outside of 23°C ±10°C (73°C ±18°F) Operating Temperature
  • Range: -25 to 60°C (-10 to 140°F)
  • Relative Humidity Range: 10% ≤RH ≤90% (0 to 35°C), non-condensing, 10% ≤RH ≤70% (35 to 60°C), non-condensing
  • Dimensions: 143 L x 76 W x 41 mmH (5.63 x 3.00 x 1.60″)
  • Weight: 0.34 kg (12.1 oz) including boot and batteries
  • Power: 4 “AA” Alkaline 1.5V (LR6) (included)
  • Battery Life: 50 Hours low Battery low battery indication with nominal 1 hour of operation left. Protection against misconnection over-voltage protection to 60 Vdc (rated for 30 seconds)
  • Display: High contrast graphic liquid crystal display. LED backlighting for use in low lit areas.

Indicate

  • Excitation: Current 0.5 mA nominal
  • Normal Mode Rejection: 50/60 Hz, 50 dB
  • Common Mode Rejection: 50/60 Hz, 120 dB

Source

  • Accuracy: From 1 to 10.2 mA external excitation Current ± (0.015% of Reading + 0.05 Ω)
  • Below: 1 mA of external excitation
  • Current: ± (0.015% of reading + 0.025 mV/mA excitation current + 0.05 Ω)
  • Resistance Ranges: 0.00 to 410.00, 410.1 to 4001.0 Ω
  • Allowable Excitation Current Range <410 Ωs: 10.2 mA max; steady or pulsed/intermittent
  • Allowable Excitation Current Range 410 to 4001 Ωs: 1 mA max; steady or pulsed/intermittent
  • Pulsed Excitation Current compatibility: DC to 0.01 second pulse width 

Would you like to know more?

Our team can help you with any product information, quotes, or any general questions you may have about the product.

Product logos in a grid

Your no.1 solution to all of your workwear needs

Putting the ‘direct’ in directory since 2009. With our global catalogue of quality safety and environmental monitoring products all under one roof, we put an end to lengthy, saturated searches. At OSE, just find and enquire and we will put you in touch with the right person to make your product purchase.

Product logos in a grid

Product Selected

Universal RTD Calibrator

Make an Enquiry

"*" indicates required fields

*
Country
Hidden
This field is for validation purposes and should be left unchanged.

In a rush? Call our team today

|

Get the newsletter - Keep up to date with Health, Safety & Environmental monitoring products from across the globe with our free newsletter