High Accuracy Digital RTD Thermometer & Data Logger

DP9601
Download PDFs:
PDF Icon

The DP9601 provides precise platinum resistance thermometer temperature measurement and logging for exacting applications in both laboratory and site locations. It is an exceptionally accurate and stable digital thermometer particularly suitable for use with thermal calibration baths as a reference standard. Applications include medical, pharmaceutical, food, environmental testing, R&D and educational establishments. Displayed values and user information are indicated on a bright clear LED screen with diffused backlighting. Data can be displayed in °C, °F, K, and ohms as required; nine front panel push keys, the only user controls, are used in conjunction with the display screen. It is this arrangement which makes for very simple and “friendly” operation.

There are two input ports for 3- or 4-wire Pt100 sensors; the instrument automatically recognizes 3- or 4-wire configurations. The measured temperature can be displayed directly from one of the inputs or differentially between the two inputs. Differential temperature and the two individual channel temperatures are displayed simultaneously. A flash drive USB port is provided to allow data to be stored and/or exported. PC software running in Windows® is provided as standard; it allows programming of custom calibration, remote control and measure and logging functions.

Specifications Inputs/Ranges/Sensors

  • Type: Pt100 to IEC 751 (ITS 90 refers) -200 to 850°C Ro = 100 O 3- or 4-wire connection with automatic recognition (with manual override)
  • Overall Accuracy: ±0.01°C ±0.0005% of span
  • Linearisation Conformity: Better than ±0.01°C Stability (vs Ambient Temperature): Better than 0.0025°C per 1°C ambient change
  • Warm-Up: Negligible under normal ambient conditions. Allow 5 to 10 minutes for full stability unless stored at low temperature, then 30 minutes minimum
  • Pt100 Sensor Current: 0.5 mA
  • Resolution of Data Display: 0.01
  • Variable Filter: Sampling rate selectable between 4 and 64 (measurements averaged per reading)
  • Measurement Units: °C, °F, K, Ω
  • Measurement Modes: A, B or A-B
  • Custom Calibration: Up to 10 calibration values can be allocated to Pt100 input A & B and to thermocouples input A & B; values are retained in non-volatile memory until replaced by user; ITS 90 or IPTS 68 coefficients can be used for custom calibration
  • Smart Sensor Connection: Correction values stored in connector
  • Null Function: Corrects differential temperature readout between two sensors to zero
  • Sensor Lead Resistance: 5 Ω each lead maximum
  • Logging: 8000 readings
  • Supply: Internal lithium Ion rechargeable batteries; mains 90 to 260V 50/60 Hz universal adaptor included; battery charge life up to 12 hours with full charge.
  • Series Mode Rejection: 60 dB @ 50 Hz (50 mV RMS applied)
  • Common Mode Rejection: 30V RMS applied between input and earth produces no measurable effect
  • Display: 128 x 64 pixel character OLED screen with diffused backlighting; wide viewing angle, high contrast (adjustable)
  • Front Panel Controls: 9 membrane push-keys to control all instrument functions
  • Mechanical/Case: Metal bench top case/adjustable tilt
  • Dimensions: Overall 190 W x 70 H x 250 mm D 
  • Weight: 3 kg approx
  • Input Connections: 2 x Pt100 via D type connectors
  • USB Serial: Isolated, 38400 Baud, 8 data, no parity, 1 stop bit
  • Communications: Remote control and measure
  • PC Software (Standard): Supplied as standard on CD-ROM. Remote control and measure: log readings to file/ download to PC/programming corrections
  • Analogue Output (Standard): User programmable, 0 to 1 V
  • Application Note: Inputs are not isolated in the instrument, which is primarily designed for laboratory applications and site calibration of industrial temperature sensors. Probes connected to the instrumentmust therefore be isolated from high voltage pick-up.
  • Precision Pt100 Probes: 304 Stainless steel probes, 6 mm Dia. with 2 m screened PTFE lead Pt100 Probes and connector
  • DP9600-L250: 250 mm L, -50 to 250°C 
  • DP9600-H450: 350 mm L, -50 to 450°C 

The DP9601 provides precise platinum resistance thermometer temperature measurement and logging for exacting applications in both laboratory and site locations. It is an exceptionally accurate and stable digital thermometer particularly suitable for use with thermal calibration baths as a reference standard. Applications include medical, pharmaceutical, food, environmental testing, R&D and educational establishments. Displayed values and user information are indicated on a bright clear LED screen with diffused backlighting. Data can be displayed in °C, °F, K, and ohms as required; nine front panel push keys, the only user controls, are used in conjunction with the display screen. It is this arrangement which makes for very simple and “friendly” operation.

There are two input ports for 3- or 4-wire Pt100 sensors; the instrument automatically recognizes 3- or 4-wire configurations. The measured temperature can be displayed directly from one of the inputs or differentially between the two inputs. Differential temperature and the two individual channel temperatures are displayed simultaneously. A flash drive USB port is provided to allow data to be stored and/or exported. PC software running in Windows® is provided as standard; it allows programming of custom calibration, remote control and measure and logging functions.

Specifications Inputs/Ranges/Sensors

  • Type: Pt100 to IEC 751 (ITS 90 refers) -200 to 850°C Ro = 100 O 3- or 4-wire connection with automatic recognition (with manual override)
  • Overall Accuracy: ±0.01°C ±0.0005% of span
  • Linearisation Conformity: Better than ±0.01°C Stability (vs Ambient Temperature): Better than 0.0025°C per 1°C ambient change
  • Warm-Up: Negligible under normal ambient conditions. Allow 5 to 10 minutes for full stability unless stored at low temperature, then 30 minutes minimum
  • Pt100 Sensor Current: 0.5 mA
  • Resolution of Data Display: 0.01
  • Variable Filter: Sampling rate selectable between 4 and 64 (measurements averaged per reading)
  • Measurement Units: °C, °F, K, Ω
  • Measurement Modes: A, B or A-B
  • Custom Calibration: Up to 10 calibration values can be allocated to Pt100 input A & B and to thermocouples input A & B; values are retained in non-volatile memory until replaced by user; ITS 90 or IPTS 68 coefficients can be used for custom calibration
  • Smart Sensor Connection: Correction values stored in connector
  • Null Function: Corrects differential temperature readout between two sensors to zero
  • Sensor Lead Resistance: 5 Ω each lead maximum
  • Logging: 8000 readings
  • Supply: Internal lithium Ion rechargeable batteries; mains 90 to 260V 50/60 Hz universal adaptor included; battery charge life up to 12 hours with full charge.
  • Series Mode Rejection: 60 dB @ 50 Hz (50 mV RMS applied)
  • Common Mode Rejection: 30V RMS applied between input and earth produces no measurable effect
  • Display: 128 x 64 pixel character OLED screen with diffused backlighting; wide viewing angle, high contrast (adjustable)
  • Front Panel Controls: 9 membrane push-keys to control all instrument functions
  • Mechanical/Case: Metal bench top case/adjustable tilt
  • Dimensions: Overall 190 W x 70 H x 250 mm D 
  • Weight: 3 kg approx
  • Input Connections: 2 x Pt100 via D type connectors
  • USB Serial: Isolated, 38400 Baud, 8 data, no parity, 1 stop bit
  • Communications: Remote control and measure
  • PC Software (Standard): Supplied as standard on CD-ROM. Remote control and measure: log readings to file/ download to PC/programming corrections
  • Analogue Output (Standard): User programmable, 0 to 1 V
  • Application Note: Inputs are not isolated in the instrument, which is primarily designed for laboratory applications and site calibration of industrial temperature sensors. Probes connected to the instrumentmust therefore be isolated from high voltage pick-up.
  • Precision Pt100 Probes: 304 Stainless steel probes, 6 mm Dia. with 2 m screened PTFE lead Pt100 Probes and connector
  • DP9600-L250: 250 mm L, -50 to 250°C 
  • DP9600-H450: 350 mm L, -50 to 450°C 

Would you like to know more?

Our team can help you with any product information, quotes, or any general questions you may have about the product.

Product logos in a grid

Your no.1 solution to all of your workwear needs

Putting the ‘direct’ in directory since 2009. With our global catalogue of quality safety and environmental monitoring products all under one roof, we put an end to lengthy, saturated searches. At OSE, just find and enquire and we will put you in touch with the right person to make your product purchase.

Product logos in a grid

Product Selected

High Accuracy Digital RTD Thermometer & Data Logger

Make an Enquiry

"*" indicates required fields

*
Country
Hidden
This field is for validation purposes and should be left unchanged.

In a rush? Call our team today

|

Get the newsletter - Keep up to date with Health, Safety & Environmental monitoring products from across the globe with our free newsletter