EE680 – Air Velocity Sensor for Laminar Flow Monitoring

Download PDFs:
PDF Icon

The EE680 air velocity sensor by E+E Elektronik allows to monitor the laminar flow in cleanrooms or on safety workbenches. The sensor precisely measures the air velocity from 0.1 to 2 m/s (20…400 ft/min) and simultaneously also the temperature. It meets the requirements of Good Manufacturing Practice (GMP) and is therefore ideally suited for the pharmaceuticals, biotechnology and microelectronics industries.

High Accuracy and Long-Term Stability

Factory multi-point adjustment of the air velocity ensures high measurement accuracy over the entire working range. An adjustment point at 0.45 m/s (88.6 ft/min) takes into account the air flow range defined in the EU GMP directive for the manufacture of sterile medical devices. The measurement accuracy of the sensor is confirmed by an inspection certificate in accordance with DIN EN 10204-3.1.

The innovative flow sensing element used in the EE680 is based on the hot film anemometer principle and offers excellent long-term stability and low angular dependence. The special E+E sensor coating makes the thin-film sensing element highly resistant to H2O2 sterilisation and other aggressive cleaning agents.

GMP-Compliant Design

With its smooth stainless steel enclosure and a sensor design optimised for laminar flow monitoring, the EE680 meets the requirements of Good Manufacturing Practice (GMP) in the cleanroom. The straight or angled design of the sensor with different probe lengths allows for wall or ceiling mounting at the ideal distance from the filter element, according to EN ISO 14644-3. The practical mounting flange and M12 stainless steel connector facilitate the installation and replacement of the EE680.

Visual Status Indication

An LED ring integrated into the enclosure indicates the laminar flow conditions and the sensor status. Any deviation from the target state or operating fault can therefore be quickly detected. The preset threshold values and further indication options can be configured individually.

Analogue or Digital Output The measured values for air velocity and temperature are optionally available at the analogue voltage or current output, or at the RS485 interface via the Modbus RTU protocol. The type of measured value output can also be retroactively configured.

The EE680 air velocity sensor by E+E Elektronik allows to monitor the laminar flow in cleanrooms or on safety workbenches. The sensor precisely measures the air velocity from 0.1 to 2 m/s (20…400 ft/min) and simultaneously also the temperature. It meets the requirements of Good Manufacturing Practice (GMP) and is therefore ideally suited for the pharmaceuticals, biotechnology and microelectronics industries.

High Accuracy and Long-Term Stability

Factory multi-point adjustment of the air velocity ensures high measurement accuracy over the entire working range. An adjustment point at 0.45 m/s (88.6 ft/min) takes into account the air flow range defined in the EU GMP directive for the manufacture of sterile medical devices. The measurement accuracy of the sensor is confirmed by an inspection certificate in accordance with DIN EN 10204-3.1.

The innovative flow sensing element used in the EE680 is based on the hot film anemometer principle and offers excellent long-term stability and low angular dependence. The special E+E sensor coating makes the thin-film sensing element highly resistant to H2O2 sterilisation and other aggressive cleaning agents.

GMP-Compliant Design

With its smooth stainless steel enclosure and a sensor design optimised for laminar flow monitoring, the EE680 meets the requirements of Good Manufacturing Practice (GMP) in the cleanroom. The straight or angled design of the sensor with different probe lengths allows for wall or ceiling mounting at the ideal distance from the filter element, according to EN ISO 14644-3. The practical mounting flange and M12 stainless steel connector facilitate the installation and replacement of the EE680.

Visual Status Indication

An LED ring integrated into the enclosure indicates the laminar flow conditions and the sensor status. Any deviation from the target state or operating fault can therefore be quickly detected. The preset threshold values and further indication options can be configured individually.

Analogue or Digital Output The measured values for air velocity and temperature are optionally available at the analogue voltage or current output, or at the RS485 interface via the Modbus RTU protocol. The type of measured value output can also be retroactively configured.

Would you like to know more?

Our team can help you with any product information, quotes, or any general questions you may have about the product.

Product logos in a grid

Your no.1 solution to all of your workwear needs

Putting the ‘direct’ in directory since 2009. With our global catalogue of quality safety and environmental monitoring products all under one roof, we put an end to lengthy, saturated searches. At OSE, just find and enquire and we will put you in touch with the right person to make your product purchase.

Product logos in a grid

Product Selected

EE680 – Air Velocity Sensor for Laminar Flow Monitoring

Make an Enquiry

"*" indicates required fields

*
Country
Hidden
This field is for validation purposes and should be left unchanged.

In a rush? Call our team today

|

Get the newsletter - Keep up to date with Health, Safety & Environmental monitoring products from across the globe with our free newsletter