Flange Mount Zirconium Dioxide Oxygen Sensor Wired

The Flange Mount Zirconium Dioxide Oxygen Sensor is designed for easy sealing into a manifold or sampling chamber. Favoured by SST’s aerospace customers and in applications that require high temperature and pressure operation.

SST’s range of zirconium dioxide oxygen sensors use a unique closed loop measurement system. While the sensor continuously measures the oxygen level, a heartbeat signal is also generated providing an immediate warning of any failures.

Their long operating life is due to the non-depleting sensor cell technology. No reference gas is required and easy single point calibration can be performed in any known gas including fresh air.

The Flange Mount oxygen sensor offers complete with 0.4m of 5 single strand 20AWG cables and MOLEX 39-01-2060 connector, a temperature range of -100 to +250⁰C and response time of less than 4 seconds. The oxygen sensor offers a linear output signal, a lifetime of up to 10 years (depending on the application) and covers an oxygen range 0.1 to 100% O₂.

All of SST’s zirconia oxygen sensors require interface electronics to operate which SST can provide (see related products). For high volume applications, we can work with you to design your own electronics, saving you time and money.

TECHNICAL SPECIFICATIONS
Heater voltage
Operating 4.35VDC ± 0.1VDC (1.85A)
Standby 2VDCDC (0.85A)
Pump impedance at 700°C < 6kΩ
Permissible gas temperature -100°C to +250°C
Gas flow rate 0—10 m/s
Repetitive permissible acceleration 5g
Incidental permissible acceleration 30g
Termination 0.4m cable with Molex connector

OUTPUT VALUES
Oxygen pressure range 2mbar—3bar max.
Accuracy 5mbar max
Internal operational temp 700°C
Response time (10—90% step) < 4s
Warm up time (prior to sensor operation) 100s
Warm up time (from standby) 20s
Output stabilisation time ~ 180s

 

Miniature Zirconium Dioxide Oxygen Sensors

The Miniature Zirconium Dioxide Oxygen Sensors can be PCB mounted and are used when measuring ambient gas with the complete sensor immersed in the atmosphere which is to be measured.

SST’s range of Miniature Zirconium Dioxide Oxygen Sensors use a unique closed loop measurement system. While the sensor continuously measures the oxygen level, a heartbeat signal is also generated providing an immediate warning of any failures.

Their long operating life is due to the non-depleting sensor cell technology. No reference gas is required and easy single point calibration can be performed in any known gas including fresh air.

The Miniature oxygen sensor is just 12mm in diameter making it suitable for applications where space is limited. The sensor offers an option of two response times and a temperature range of -100 to 250⁰C. The oxygen sensor offers a linear output signal, a lifetime of up to 10 years (depending on the application), and covers an oxygen range from 0.1 to 100% O₂.

All of SST’s zirconia oxygen sensors require interface electronics to operate which SST can provide (see related products below). For high volume applications, we can work with you to design your own electronics, saving you time and money.

TECHNICAL SPECIFICATIONS
Heater voltage
Standard response sensor 4VDC ± 0.1VDC (1.7A)
Standby 1.65VDC (0.7A)
Fast response sensor 4.35VDC ± 0.1VDC (1.85A)
Standby 2VDC (0.85A)
Pump impedance at 700°C < 6kΩ
Permissible gas temperature 100°C to +250°C
Gas flow rate 0—10 m/s
Repetitive permissible acceleration 5g
Incidental permissible acceleration 30g

OUTPUT VALUES
Oxygen pressure range 2mbar—3bar max
Accuracy 5mbar max
Internal operational temperature 700°C
Response time (10—90% step)
Standard response sensor <15s
Fast response sensor < 4s Warm
Warm up time (prior to sensor operation) 60s
Warm up time (from standby) 20s
Output stabilisation time ~ 180s

O2I-FLEX Oxygen Sensor Interface Electronics

The O2I-Flex Oxygen Sensor Interface Electronics provides the electronics necessary to power and control SST’s range of zirconium dioxide oxygen sensors.

The O2I-Flex oxygen sensor interface electronics can be user configured to output measuring ranges of 0-25% O₂ and 0-100% O₂. The entire measurement range is linear in both cases and offers three output options 0-10V, 4-20mA and RS232 comms interface.

The O2I-Flex can be calibrated in fresh air (20.7% O₂) or in any other known O₂ concentration by either an automatic or manual external trigger or via an on-board push button switch. The electronics offer power and sensor operating LEDs and removable polarised screw terminals for easy wiring.

The adaptive software filtering provides a fast sensor response coupled with a stable oxygen output.

TECHNICAL SPECIFICATIONS
Supply voltage 24VDC ± 10%
Supply current 600mA max. at 24VDC
Digital output RS232
Analogue output 4—20mA; load 100—600Ω
0—10VDC; load 10kΩ min
Temperature limits
Storage -10 to 70⁰C
Operating -10°C to +70°C
Oxygen pressure limits 1—1000mbar

OUTPUT VALUES
Oxygen range (analogue output) 0.12—25% O2
0.12—100% O2
Oxygen range (RS232 output) 0.12—100% O2
Accuracy after calibration 1% O2
Repeatability after calibration 0.5% O2
Output resolution:
0—10VDC: 0.01V
4—20mA: 0.01mA
RS232: 0.01% O2
Response time (step 10—90%)
Fast response sensor connected: < 4s
Standard response sensor connected: < 15s
Initial warm up time (till stable output) 5—10mins
Output inactive start up delay (heater warm up) 6

OXY-LC Oxygen Sensor Interface Electronics

The OXY-LC Oxygen Sensor Interface Electronics provides the necessary electronics to power and control SST’s range of zirconium dioxide (ZrO₂) sensors.

The OXY-LC Oxygen Sensor offers various user configurable output variants RS485, PWM, 0-10Vdc and 4-20mA. The RS485 Modbus RTU Protocol offers various benefits over other output types but most importantly it allows complete control of the sensor operation and access to all available information including sensor diagnostics, barometric pressure and pressure sensor temperature.

The adaptive software filtering which enables it to provide a fast response coupled with a stable oxygen output. In addition to this, the OXY-LC Oxygen sensor interface electronics offers a high accuracy linear output for all of our zirconium dioxide oxygen sensors.

The built in pressure sensor allows for barometric pressure compensation and the interface mounted screw terminals offers easy wiring with reverse voltage and transient overvoltage.

TECHNICAL SPECIFICATIONS
Supply voltage
RS485 variant: 8—28VDC
Analogue variants: 20—28VDC
Supply current 600mA max. at 24VDC
1.2A max. at 12VDC
Digital output RS485 (Modbus RTU)
PWM
Analogue output 4—20mA; load 600Ω max
0—10VDC; load 10kΩ min
Temperature limits
Storage: -40°C to +85°C
Operating: 30°C to +70°C
Operating pressure limits 260—1260mbar absolute

OUTPUT VALUES
Oxygen range (analogue output) 0.1—25% O2
0.1—100% O2
Oxygen range (Modbus RTU) 0.1—100% O2
Accuracy after calibration 0.5% O2
Repeatability after calibration 0.01% O2
Response time (step 10—90%)
Fast response sensor connected: 4s
Standard response sensor connected: 15s
Initial warm up time (till stable output) 5—10mins
Output inactive start up delay (heater warm up)
From OFF to ON mode: 60s
From Standby to ON mode (RS485 only): 20s

Screwfit Zirconium Dioxide Oxygen Sensors

SST manufacture the Screwfit zirconium dioxide oxygen sensors, which is ideal for applications in the combustion industry.

The Screwfit zirconium dioxide oxygen sensors have a threaded probe design making it easy to install. Suitable for combustion control within industrial coal/oil/gas/biomass boilers, continuous emission monitoring systems and controlled atmosphere monitoring.

SST’s range of Screwfit zirconium dioxide oxygen sensors use a unique closed loop measurement system. While the sensor continuously measures the oxygen level, a heartbeat signal is also generated providing an immediate warning of any failures.

Their long operating life is due to the non-depleting sensor cell technology. No reference gas is required and easy single point calibration can be performed in any known gas including fresh air.

The oxygen sensor can cope with extreme temperature levels from -100°C up to 250°C and has a rapid response time of less than four seconds. In addition to this, the sensor offers a linear output signal, a lifetime of up to 10 years (depending on the application), and covers an oxygen range from 0.1 to 100% O₂.

All of SST’s zirconia oxygen sensors require interface electronics to operate which SST can provide (see related products below). For high volume applications, we can work with you to design your own electronics, saving you time and money.

TECHNICAL SPECIFICATIONS
Heater voltage
Operating 4.35VDC ± 0.1VDC (1.85A)
Standby 2VDC (0.85A)
Pump impedance at 700°C3 < 6kΩ
Permissible gas temperature -100°C to +250°C
Gas flow rate 0—10 m/s
Repetitive permissible acceleration 5g
Incidental permissible acceleration 30g
Mounting thread M18 x 1.5

OUTPUT VALUES
Oxygen pressure range 2mbar—3bar max
Accuracy 5mbar max
Internal operational temp 700°C
Response time (10—90% step) < 4s
Warm up time (prior to sensor operation) 60s
Warm up time (from standby) 20s
Output stabilisation time ~ 180s

 

Flange Mount Zirconium Dioxide Oxygen Sensor Pinned

The Flange Mount Zirconium Dioxide Oxygen Sensor is designed for easy sealing into a manifold or sampling chamber. Favoured by SST’s aerospace customers and in applications that require high temperature and pressure operation.

SST’s range of zirconium dioxide oxygen sensors use a unique closed loop measurement system. While the sensor continuously measures the oxygen level, a heartbeat signal is also generated providing an immediate warning of any failures.

Their long operating life is due to the non-depleting sensor cell technology. No reference gas is required and easy single point calibration can be performed in any known gas including fresh air.

The Flange Mount oxygen sensor offers 4 or 5 pin termination, a temperature range of -100 to 400⁰C and response time of less than 4 seconds. The oxygen sensor offers a linear output signal, a lifetime of up to 10 years (depending on the application) and covers an oxygen range 0.1 to 100% O₂.

All of SST’s zirconia oxygen sensors require interface electronics to operate which SST can provide (see related products). For high volume applications, we can work with you to design your own electronics, saving you time and money.

TECHNICAL SPECIFICATIONS
Heater voltage
Operating 4.35VDCDC ± 0.1VDC (1.85A)
Standby 2VDC (0.85A)
Pump impedance at 700°C < 6kΩ
Permissible gas temperature -100°C to +400°C
Gas flow rate 0—10 m/s
Repetitive permissible acceleration 5g
Incidental permissible acceleration 30g

OUTPUT VALUES
Oxygen pressure range 2mbar—3bar max
Accuracy 5 mbar max
Internal operational temp 700°C
Response time (10—90% step) < 4s
Warm up time (prior to sensor operation) 60s
Warm up time (from standby) 20s

Probe Zirconium Dioxide Oxygen Sensors

The Probe zirconium dioxide oxygen sensors are available in three lengths. Suitable for aggressive and polluted applications such as boiler exhausts. Longer probes are an ideal solution for thicker walled, insulated flue pipes.

SST’s range of zirconium dioxide oxygen sensors use a unique closed loop measurement system. While the sensor continuously measures the oxygen level, a heartbeat signal is also generated providing an immediate warning of any failures.

Their long operating life is due to the non-depleting sensor cell technology. No reference gas is required and easy single point calibration can be performed in any known gas including fresh air.

The Probe zirconium dioxide oxygen sensors offer standard and fast response times from as little as four seconds. Various probe lengths, cable lengths, and a standard and extended temperature range are available. The Probe zirconium dioxide oxygen sensor offer a linear output signal, a lifetime of up to 10 years (depending on the application), and covers an oxygen range from 0.1 to 100% O₂.

All of SST’s zirconia oxygen sensors require interface electronics to operate which SST can provide (see related products below). For high volume applications, we can work with you to design your own electronics, saving you time and money.

TECHNICAL SPECIFICATIONS
Heater voltage
Operating (standard response) 4VDC ± 0.1VDC (1.7A)
Standby 1.65VDC (0.7A)
Operating (fast response) 4.35VDC ± 0.1VDC (1.85A)
Standby 2VDC (0.85A)
Pump impedance at 700°C < 6kΩ
Permissible gas temperature -100°C to +250°C
Gas flow rate 0—10 m/s
Repetitive permissible acceleration 5g
Incidental permissible acceleration 30g

OUTPUT VALUES
Oxygen pressure range 2mbar—3bar max
Accuracy 5mbar max
Internal operational temperature 700°C
Response time (10—90% step)
Standard response sensor < 15s
Fast response sensor < 4s
Warm up time (prior to sensor operation) 60s
Warm up time (from standby) 20s
Output stabilisation time ~ 180s

 

Subscribe

Subscribe to our newsletter now and stay up-to-date with new products collections, the latest and exclusive offers.

You must tick this box in order to submit your enquiry or survey response. By doing so you consent to this company gathering the personal data you have entered in your enquiry or survey. We will share this information with the product provider to enable them to contact you. Please refer to our privacy policy on this site for further information.